Products

ColiMinder™ CMI-02 Low Energy Version

Product details

ColiMinder® CMI-02-LE is a low-energy version of Industrial ColiMinder® CMI-02. It can be powered by 12V batteries, 100/240V to 12V ACDC power supply or by car electrical system. ColiMinder® CMI-02-LE serves in areas with reduced power supply for long-term monitoring of bio-contaminated water samples. It can be installed in areas with rough environmental conditions (IP65 protection).
Its fully automated measurement system can measure microbiological contamination of liquid samples within 15 minutes.
The ColiMinder® is monitoring microbiological contamination through the specific enzymatic activity of the respective target organism.

Its fully automated measurement system is able to measure microbiological contamination of liquid samples within 15 minutes.

The ColiMinder™ is monitoring microbiological contamination through the specific enzymatic activity of the respective target organism.

features

E

Basic Principles

Fluorometric measurement technology

Reagents & Consumables

  • Two component measurement reagents
    (CM Quick Detect)
  • Cleaning solution (CM Quick Clean)
  • Rinsing water (not provided)
  • No contaminated or hazardous waste is produced
  • Reagents holding capacity: up to 100 measurements

Key Features

R

fully automated sampling, measurement, cleaning and calibration

R

Online data visualization and automatic notification

R

100 measurements without staff intervention

R

up to 54 (80 in special cases) measurements per day

R

Fully controllable through internet connection

R

2 sample intakes (more optional)

Measurement Frequency

  • 15 min measurement time followed by a 9 min cleaning cycle
  • Continuous Mode: up to 54 measurements per day.
  • Interval Mode: time between the measurements can be set manually
  • Persistent Mode: Sample 1 or Sample 2
  • Alternating Mode: Sample 1 and Sample 2
  • The measurement process is a batch process.

For a short time, measurement frequency can be increased to approx. 80 measurements per day (with reduced cleaning).

Data transfer and visualization

Measurements data is directly transmitted to server through internet/ network connection

Live data visualization on dedicated website, results can be downloaded

Automatic notifications can be set (E-mail, SMS)

Measurement results are also available through RS232/RS485 or Modbus TCP (optional)

Measurement results can be saved on USB Flash Drive directly from the device

4 to 20 mA output optional

Remote control

All functions and programming can be performed through remote control (internet or network connection required). The ColiMinder® is able to control ISCO 6700 auto samplers, which also can be controlled remotely through ColiMinder® remote control.

Online visualization

Using this service, you are able to monitor status and results of your ColiMinder live and online on a dedicated homepage. Results are presented in a timeline and can be downloaded in csv or xls

Add-On Modules

10-fold multisampling module 50 ml bottles

sample line extension up to 10 intakes

ISCO Auto-Sampler Controller

To trigger and control ISCO 6700 auto sampler

10-fold multisampling module 250 ml bottles

sample line extension up to 10 intakes

Data Output Module 4-20 mA
Sampling Unit – sample from a stream SU-1

Necessary when sampling from a pressurized stream

Installation Rack

Tailor-made rack for installations on-site, including holdings for device, canisters, tablet

Connection Ports

WIFI

LAN

RS-232

USB

HDMI

Internal GPRS/LTE–Modem

AutoSampler Control Port

Alarm–Relay output or alarm flashlight

SpecificationS

Measurement time for quantitative results15 min.
Performance (measurements per day)Standard: up to 56 (1x cleaning) or up to 45 (2x cleaning)
Special case (with reduced cleaning): up to 80
Number of sample intakesStandard: 2 (optional - 10)
Housing:IP 65 (powder coated aluminum)
Full working cycle incl. automatic cleaning< 30 min.
Power supply12V DC
Space for device W x H x D60 cm x 60cm x 30cm
Dimensions W x H x D55 cm x 50cm x 21 cm
Weight15 kg
Detection limit (E.coli)0.8 mMFU/100ml
Detection limit (Enterococci)0.5 µGU/100ml
Detection limit (Total Bacteria)0.5 µU/100ml
Temperature range for operation0°C to 40°C
Reagent cooler0°C-20°C, 90W

Need further details? Send us a note!